Liquefied
gas LNG sample probe & vaporizer for chromatograph / calorimeter / spot or
automatic sampling automatic
sampling

For LNG
terminal or recondenser it is now a common practice to determine the
composition, density and gross calorific value of LNG delivered by direct on
line GC or wobbe analyser plus indirect lab analysis methods. A precise sampling
is prerequisite for both analysis.
For more information on
the LNG sample takeoff and vaporizer, click here
For
more information on the LNG vaporizer descriptive, click here
I- LNG Sample take-off probe:
The ISOSAMPLE 8100 SAMPLE TAKE-OFF PROBE has
been designed to provide a thermal insulation of the sample take-off probe by
hight vacuum maintained below 1.10-4 torr.
The
conjonction of this ultra-low residual pressure minimizing heat leak by
conduction and convection with layer of radiation shields cryogenic perlite
constitutes a superinsulation system providing apparent thermal conductivity (k)
values below 0.4 milliwatt per meter-kelvin (mW/m-k), therefore the enthalpy
rise can be controlled to be inferior of the subcooling degree and
fractionation will not have occured in the sampling line even for low LNG
pressure in line
II-LNG Sample
vaporizer:
The LNG sample is vacuum isolated up to the
sample vaporizer coil exchanger sealed in calorific block temperature
controlled.
At this point, sample is flashing to eliminate any fractionated distillation
effect.
Inlet / outlet temperatures and pressures are
monitored with interlock shut-off system
This principle focuses on improving
reproducibility between lab and on line GC to a value better than the
0.15% specified by ASTM D-1945 .
The vaporizer is provided
in a stainless steel box, protecting code is ATEX II 2G EEx de IICT4